Select a standard against which you will measure your readings. For example, some grasses will be best for baling between 24 and 30 percent moisture. In this instance you may choose 26 or 27 percent as your standard for calibration. Some models will default to a pre-set figure, in which case you must change the set point.
Run a calibration test on the instrument. There may be a single button for this task or you may be required to press several, either in sequence or simultaneously. Your owner's manual will guide you in this procedure.
Power off the meter and clean the probes if the display is giving illogically high or low moisture readings. It may be reading the moisture content of foreign debris. Excessive humidity will also affect calibration. Perform this task in a climate controlled environment.
Turn the meter back on and confirm that the set point has not changed. Repeat the calibration test.